- Purity
- Speed
- Resolution
- Ligand
- Ligand Density
- Available Capacity: It is the actual amount of protein that can bind to the media. If flow rate is included in the defined conditions then it is called as Dynamic Binding Capacity.
Dynamic Binding Capacities of Different Matrices |
- Salt Concentration
- Flow Rate
- Temparature
- pH to a lesser extent
Most commonly used salts in HIC are
Ammonium Sulfate, Sodium Sulfate, Sodium Chloride, Potassium Chloride, and Ammonium acetate.
Characteristics of good HIC Matrix
- High Binding Capacity – Larger Area will results in high capacity binding
- Physical Stability
- Chemical Stability
- Matrix should be Inert
Check the stability of protein at different salt concentrations.
Proteins shouldn’t precipitate at high salt concentration, if protein precipitates there will no or very low yield.
Elution from the HIC column is in the increasing order of hydrophobicity.
Ligands Used Hydrophobic Interaction Chromatography meida has Alkyl or Aryl Groups.
Phenyl 650 |
- Alkyl – Butyl, Octyl, Ether, Isopropyl – Eg: Butyl S
- Aryl – Phenyl – Eg: Phenyl 650 –
For tightly bound proteins use harsh conditions like washing with 0.5M – 1.0M NaOH, 70% Ethanol or 30% IPA followed by water washes or salt free buffer washes.
References
Encyclopedia of Bioprocess Technology: FERMENTATION,BIOCATALYSIS, & BIOSEPARATION
Technical Resources on Chromatographic Media From GE, TOSOH, Shimadzu
Got something to say about this post? Leave a comment…your comments are valuable for improving the posts.