Importance of Column Packing & Efficiency Calculations

Column Packing is one of the most important step in the chromatographic purification process. A well packed column yields higher resolution and yield.
Column Packing Efficiency can be checked by calculating:
  • HETP – Height Equivalent Theoretical Plates
  • Peak Asymmetry Factor (AF / As)
HETP Calculation:
HETP = ( d2 * L ) / (5.54* t2)

Where: t = retention value expressed as units of time, volume or measured distance on the Chromatogram
d = peak width at half height expressed in same units as “t”
L = column length (in units of cm for HETP or m for N/m)

N/m =  (5.54* t2) / ( d2 * L )

Asymmetry Factor (AF / As) Calculation:


chromatography peak asymmetry

AF = b/a

Where: a is the peak width (left half) at 10% of peak height, b is the peak width (right half) at 10% of peak height, 


Column Packing Guidelines – TOSOH Biosciences 

References
TOSOH Biosciences – Technical Resources
GE Lifescience – Technical Resources
Pall Lifesciences – Technical Resources

Got something to say about this post? Leave a comment…your comments are valuable for improving the posts.